Calibrator of new lot how to feed in fully automatic biochemistry analyzer xl640
Автор: Manoj MLT
Загружено: 2026-02-02
Просмотров: 42
Описание:
how to feed new lot of calibrator in fully automatic biochemistry analyzer
1️⃣ Check the calibrator first
Confirm lot number, expiry date, and storage condition
Reconstitute (if lyophilized) exactly as per the insert
Mix gently (❌ no vigorous shaking)
2️⃣ Enter calibrator lot details in the analyzer
Go to: Calibration / Reagent / QC / Utility menu (name depends on analyzer)
Enter:
✅ Calibrator name
✅ New lot number
✅ Expiry date
✅ Assigned values (if required manually)
✅ Factor / concentration (if mentioned in insert)
📌 Some analyzers auto-read this via barcode; others need manual entry.
3️⃣ Load the calibrator on the analyzer
Place calibrator vial/cup in the specified position
Select correct calibrator level (Cal 1, Cal 2, multi-point, etc.)
Ensure enough volume is loaded
4️⃣ Perform calibration
Select tests to be calibrated
Start calibration run
Wait for completion
Analyzer will show:
✔️ Calibration Accepted
or
❌ Calibration Failed
5️⃣ Review calibration status
Check:
Calibration curve
Slope / intercept
Calibration flags or error messages
⚠️ If failed:
Recheck lot entry
Re-mix calibrator
Clean probe / cuvette
Repeat calibration
6️⃣ Run QC after calibration (VERY IMPORTANT)
Run both normal & pathological QC
QC must be within range
Only then release patient reports
Common mistakes to avoid ❌
Using old calibration data with new lot
Skipping QC after calibration
Wrong assigned value entry
Expired or improperly stored calibratorj
#howtofeedcalibratornewlotconcentrTionvalue
#howyocalibratfullyautomayicbiochemistryanalyzer
#xl640calibrationnewlot
#howtocalibratbiochemistrymanalyzer