ycliper

Популярное

Музыка Кино и Анимация Автомобили Животные Спорт Путешествия Игры Юмор

Интересные видео

2025 Сериалы Трейлеры Новости Как сделать Видеоуроки Diy своими руками

Топ запросов

смотреть а4 schoolboy runaway турецкий сериал смотреть мультфильмы эдисон
Скачать

Bubble Point Test Demo by Sterlitech

Автор: Sterlitech Corporation

Загружено: 2019-01-31

Просмотров: 21853

Описание: Watch our demo on how to test for a membrane filter's bubble point, which is typically performed after flow rate has been tested. Bubble point, the point at which air pushes through the pores of a wetted membrane, relates to the retention characteristics (and pore size) of that membrane.

0:02 Prepare tester
0:40 Attach wetted filter
1:01 Turn up pressure
1:58 Increase pressure
2:29 Monitor bubble point
3:28 Remove filter
3:48 Close down tester

Sterlitech Membrane Disc Filters

Sterlitech offers a comprehensive selection of polymeric and inorganic membrane filter materials, enabling users to perform everything from solvent filtration to gas venting.

Our pre-cut membrane disc filters are produced using the highest quality manufacturing standards. These filters are designed to increase efficiency, reduce cost, and expand the capabilities of microfiltration applications.

Available membrane disc filters:

MCE Membranes - High protein binding, high flow rate
Silver Membranes - Inorganic, chemical and temperature resistant
Aluminum Oxide Membranes - Inorganic, transparent when wet
Cellulose Acetate Membranes - Low protein binding, high throughput, low extractables
Ceramic Membranes - Inorganic, chemical and temperature resistant
Glass Fiber Filters - High flow rate, high purity, high operating temperatures
Nylon Membranes - Superior strength, high protein binding, chemically compatible
PAN (Polyacrylonitrile) Membranes - Fast flow at low pressures, 6-log bacterial load reduction
PCTE (Polycarbonate) Membranes - Translucent, flat/smooth surface, low non-specific binding. Also available in black, gold, gray and PVP-Free.
PES (Polyethersulfone) Membranes - Low protein binding, durable, uniform results, high throughput, fast flow rate
PETE (Polyester) Membranes - Hydrophilic/Hydrophobic, track-etched pores, flat/smooth surface, solvent resistant, Gold Coated
PE (Polyethylene) Membranes - Highly hydrophobic, excellent chemical resistance, compatible with gamma irradiation sterilization
Polypropylene Nominal Prefilters - Composed of polypropylene fibers with nominal pore size rating
PTFE Membranes - Chemical and pH resistance, laminated or unlaminated
PVDF Hydrophilic Membranes - Hydrophilic, low extractable/protein binding

===========

We'll disconnect the pressure pot. We'll make sure that everything reads zero so we're safe. I think it is a good idea to open the lid just to make sure that there is no pressure in there. Then we disconnect the pressure line.

By pulling on the flange towards the fitting and then pulling the tube out, we can safely move the pressure pot out of the way.

We'll attach the test fitting back to the end of the outlet tubing. The bubble point tester outlet tubing.

Take our previously flow tested filter. We'll make sure the outlet fitting is facing upward. And we'll attach the wetted filter by screwing it clockwise until it's firm.

We'll turn up the pressure to 50% of the bubble point for a 0.45 mixed cellulose ester membrane. The particular filter that we're testing now, that would probably be the bubble point would probably be somewhere around high twenties to about 30. So we'll turn it up to 15 psi.

First we'll turn the switch on. Then we'll turn this up to 15 psi.

We'll allow that to sit for a moment, approximately 15 seconds.

After it sits for 15 seconds, we'll increase the pressure at approximately a psi minute, which is about a quarter turn a second with the handle.

We'll dab away any water that's in the way on the outside of the filter. And then we'll watch the base of this outlet stem for a bubble to appear and observe that the water will accumulate on the outlet. That's expected. And it's not the bubble point. A bit higher than I would have guessed.

There we go. Did you see the bubble? So that would be the bubble point of the filter, and it's about 48.5 psi. Checking on the gauge, you would now record the bubble point and then release the pressure by turning counterclockwise fairly rapidly until the pressure reads zero.

Now it's safe to remove the filter from the outlet fitting by screwing counterclockwise. You have to hold the fitting while you untwist it. And this filter is now tested. We'll save that if we want to do any further examination to it.

Once the bubble point testing is complete, once you're completed testing all of your filters, you make sure that you have it turned out to read zero. When you turn the gauge off, turn the air supply off, turn the gauge off, and then you're done.

===========

Interested in Sterlitech membrane filters? Find the right one to meet your application requirements here: https://tinyurl.com/224935zj

For more information, contact us here: https://tinyurl.com/4p4t4d9e or call us at 1-877-544-4420

#bubblepointtest #bubblepointtesting #membranediscfilter #membranefilter
#membranefiltration #filtration #microfiltration #sterlitech #bestpractice #demo #tutorial

Не удается загрузить Youtube-плеер. Проверьте блокировку Youtube в вашей сети.
Повторяем попытку...
Bubble Point Test Demo by Sterlitech

Поделиться в:

Доступные форматы для скачивания:

Скачать видео

  • Информация по загрузке:

Скачать аудио

Похожие видео

Introduction to Microfiltration Membrane | Webinar by Sterlitech

Introduction to Microfiltration Membrane | Webinar by Sterlitech

Wine Filtration: Integrity Testing of Membranes

Wine Filtration: Integrity Testing of Membranes

Bubble Point Testing, Membrane Disc Filters Testing by Sterlitech

Bubble Point Testing, Membrane Disc Filters Testing by Sterlitech

Filter Integrity Testing: Filter Attributes

Filter Integrity Testing: Filter Attributes

How to Perform a Visual Bubble Point on Sartolab® P20 Pressure Filtration Devices

How to Perform a Visual Bubble Point on Sartolab® P20 Pressure Filtration Devices

Мне 73. Я жалею, что понял это только сейчас.

Мне 73. Я жалею, что понял это только сейчас.

Filtering Integrity Test

Filtering Integrity Test

Я Построил Рогатку Более Мощную, чем Пистолет

Я Построил Рогатку Более Мощную, чем Пистолет

Почему все НЕПРАВИЛЬНО СВЕРЛЯТ дерево?

Почему все НЕПРАВИЛЬНО СВЕРЛЯТ дерево?

Lunch and Learn Series: Sterile Filtration/bubble point testing

Lunch and Learn Series: Sterile Filtration/bubble point testing

Понимание GD&T

Понимание GD&T

Water Entry Pressure Test Demo by Sterlitech

Water Entry Pressure Test Demo by Sterlitech

How to Assemble an Overhead Stirrer with the Sterlitech HP4750X

How to Assemble an Overhead Stirrer with the Sterlitech HP4750X

Воссоздание старинного насоса (без движущихся частей)

Воссоздание старинного насоса (без движущихся частей)

Integrity Testing Hydrophobic Filters Best Practices

Integrity Testing Hydrophobic Filters Best Practices

10 кВт! Бесплатная энергия с помощью двух электродвигателей 💡⚡💡 Liberty Engine 5.0

10 кВт! Бесплатная энергия с помощью двух электродвигателей 💡⚡💡 Liberty Engine 5.0

PDA Israel FREE Webinar :Pre Use/ Post Sterilization Integrity Test PUPSIT

PDA Israel FREE Webinar :Pre Use/ Post Sterilization Integrity Test PUPSIT

Filter Integrity Testing: Real Failure Investigations

Filter Integrity Testing: Real Failure Investigations

20 КРУТЕЙШИХ ГАДЖЕТОВ С АМАЗОНА, КОТОРЫЕ ВЫ ЗАХОТИТЕ КУПИТЬ

20 КРУТЕЙШИХ ГАДЖЕТОВ С АМАЗОНА, КОТОРЫЕ ВЫ ЗАХОТИТЕ КУПИТЬ

Невероятные свойства композитных материалов

Невероятные свойства композитных материалов

© 2025 ycliper. Все права защищены.



  • Контакты
  • О нас
  • Политика конфиденциальности



Контакты для правообладателей: [email protected]