Tutorial 11 Pressure Support Part 4: Cycling to Expiration (Esens)
Автор: CCM TUTORIALS
Загружено: 2023-04-19
Просмотров: 333
Описание:
This is the fourth and final tutorial on Pressure Support Ventilation. As you may recall, PS is an unusual mode of ventilation because it is flow cycled - that is - the ventilator cycles to expiration at a specific, user set, percentage of peak flow. The default expiratory sensitivity is usually around 25%. Expiratory dys-synchrony is frequently missed by bedside clinicians who have not been schooled in waveform analysis. This tutorial covers everything you need to know.
@ccmtutorials www.ccmtutorials.org
Contents:
00:00 Beginning and Introduction
00:40 Review of Previous Tutorial
00:55 Overview of This Tutorial
01:40 Pressure Support and Control of Flow Cycling (different ventilators)
02:20 Expiratory Sensitivity - what is it?
02:20 The Patient Continues to Inspire While The Ventilator Cycles Off
04:40 Waveform of Expiratory Dyssynchrony when flow stops too early
05:50 The Patient Expires While The Ventilator Is Still Delivering A Breath
06:10 Waveform Where Flow Continues During Exhalation
07:10 Managing Expiratory Dys-Synchrony
08:32 Clinical Scenario - Liam Recovering from ARDS
09:17 Liam Waveform
11:00 Volume Time Curves
12:00 Liam Resolution
12:40 Clinical Scenario - Brian COPD
13:20 Brian Waveform
14:50 Brian Resolution
15:35 Review of this Tutorial
16:30 Preview of Next Tutorial Series - Hypoxia